Abstract
BACKGROUND AND PURPOSE
After publications on the effectiveness of mechanical thrombectomy by stent retrievers in acute ischemic stroke with large-vessel occlusion, alternative endovascular approaches have been proposed using first-line aspiration catheters. Several devices are currently available to perform A Direct Aspiration First Pass Technique. The Sofia catheter aspiration has been widely used by interventionalists, but data are scarce about its efficacy and safety. Our aim was to report our multicenter thrombectomy experience with first-line Sofia catheter aspiration and to identify independent prognostic factors of clinical and procedural outcomes.
MATERIALS AND METHODS
We performed a retrospective analysis of the prospectively maintained Endovascular Treatment of Ischemic Stroke multicentric registry. Data from consecutive patients who benefited from thrombectomy with a first-line Sofia approach between January 2013 and April 2018 were studied. We excluded other first-line approaches (stent retriever or combined aspiration and stent retriever) and extracranial occlusions. Baseline characteristics, procedural data, and angiographic and clinical outcomes were analyzed.
RESULTS
During the study period, 296 patients were treated. Mean age and initial NIHSS score were, respectively, 69.5 years and 16. Successful reperfusion, defined by the modified TICI 2b/3, was obtained in 86.1% (n = 255; 95% CI, 81.7%–89.9%). Complete reperfusion (modified TICI 3) was obtained in 41.2% (n = 122; 95% CI, 35.5%–47.1%). A first-pass effect was achieved in 24.2% (n = 71; 95% CI, 19.4%–29.6%). A rescue stent retriever approach was required in 29.7% (n = 88; 95% CI, 24.6%–35.3%). The complication rate was 9.5% (n = 28; 95% CI, 6.4%–13.5%). Forty-three percent (n = 122; 95% CI, 37.1%–48.9%) of patients presented with a favorable 3-month outcome (mRS ≤ 2). Older age, M1-occlusion topography, and intravenous thrombolysis use prior to thrombectomy were independent predictors of the first-pass effect.
CONCLUSIONS
The first-line contact aspiration approach appeared safe and efficient with Sofia catheters. These devices achieved very high reperfusion rates with a low requirement for stent retriever rescue therapy, especially for M1 occlusions.
Read this article: http://bit.ly/2MsKuhD
Fellows’ Journal Club
Sixty-four patients with 68 saccular unruptured intracranial aneurysms were recruited. Patients underwent 3T MR imaging with 3D-TOF-MRA, 3D black-blood MR imaging, and contrast-enhanced MRA, and they underwent 3D rotational angiography within 2 weeks. The neck, width, and height of the unruptured intracranial aneurysms were measured by 2 radiologists independently on 3D rotational angiography and 3 MR imaging sequences. 3D black-blood MR imaging demonstrates the best agreement with DSA, with the smallest limits of agreement and measurement error. 3D-TOF-MRA had the largest limits of agreement and measurement error. The authors conclude that 3D black-blood MR imaging achieves better accuracy for aneurysm size measurements compared with 3D-TOF, using 3D rotational angiography as a criterion standard.