Abstract
BACKGROUND AND PURPOSE
The residual blood flow artifact is a critical confounder for MR black-blood thrombus imaging of cerebral venous sinus thrombosis. This study aimed to conduct a validation of a new MR black-blood thrombus imaging technique with enhanced blood signal suppression.
MATERIALS AND METHODS
Twenty-six participants (13 volunteers and 13 patients) underwent conventional imaging methods followed by 2 randomized black-blood thrombus imaging scans, with a preoptimized delay alternating with nutation for tailored excitation (DANTE) preparation switched on and off, respectively. The signal intensity of residual blood, thrombus, brain parenchyma, normal lumen, and noise on black-blood thrombus images were measured. The thrombus volume, SNR of residual blood, and contrast-to-noise ratio for residual blood versus normal lumen, thrombus versus residual blood, and brain parenchyma versus normal lumen were compared between the 2 black-blood thrombus imaging techniques. Segmental diagnosis of venous sinus thrombosis was evaluated for each black-blood thrombus imaging technique using a combination of conventional imaging techniques as a reference.
RESULTS
In the volunteer group, the SNR of residual blood (11.3 ± 2.9 versus 54.0 ± 23.4, P < .001) and residual blood-to-normal lumen contrast-to-noise ratio (7.5 ± 3.4 versus 49.2 ± 23.3, P < .001) were significantly reduced using the DANTE preparation. In the patient group, the SNR of residual blood (16.4 ± 8.0 versus 75.0 ± 35.1, P = .002) and residual blood-to-normal lumen contrast-to-noise ratio (12.4 ± 7.8 versus 68.8 ± 35.4, P = .002) were also significantly lower on DANTE-prepared black-blood thrombus imaging. The new black-blood thrombus imaging technique provided higher thrombus-to-residual blood contrast-to-noise ratio, significantly lower thrombus volume, and substantially improved diagnostic specificity and agreement with conventional imaging methods.
CONCLUSIONS
DANTE-prepared black-blood thrombus imaging is a reliable MR imaging technique for diagnosing cerebral venous sinus thrombosis.
Read this article: http://bit.ly/327GXZD
Fellows’ Journal Club
Twenty-six participants underwent conventional imaging methods followed by 2 randomized black-blood thrombus imaging scans, with a preoptimized DANTE preparation switched on and off, respectively. The signal intensity of residual blood, thrombus, brain parenchyma, normal lumen, and noise on black-blood thrombus images were measured. The thrombus volume, SNR of residual blood, and contrast-to-noise ratio for residual blood versus normal lumen, thrombus versus residual blood, and brain parenchyma versus normal lumen were compared between the 2 black-blood thrombus imaging techniques. The new black-blood thrombus imaging technique provided higher thrombus-to-residual blood contrast-to-noise ratio, significantly lower thrombus volume, and substantially improved diagnostic specificity and agreement with conventional imaging methods.